Home

Documentation

Project Support

video

Command:video
Contexts:Director console or command port
Function:Acquire sequential raw frames and deliver them to the video server
Required Parameters:none.
Optional Parameters:
nframes=[16] Number of raw frames/cycles
etime=*[0.005] Seconds of integration time per frame
intstart=*[0.005] Seconds after end of readout to start integration
width=[det-dependent] Width (columns) of readout per output
height=[det-dependent] Height (rows) of readout per output
videodest=[from-config] Video server destination IP, port, and feed name
keep=*[TRUE] If FALSE, read frame but discard the data
mod=Number of cycles after which to repeat pattern (must be < nframes)
ncpus=[2] The default enables multi-threaded operation for best performance. "1" disables.
dreg=*

Frame-based options

(*) Options etime= intstart= dreg= and keep= take optional extensions to change to a new value on a given frame number. For example, to perform on reset and then readouts after:

  dreg=<regval_with_reset>,2:<regval_no_reset>

Without whitespace a comma, a frame number, a colon, and new value.

nframes

The nframes option actually specifies the number of video "cycles" to complete. The video cycle goes in this order:

  • Program dreg
  • Perform a readout (possibly with integration starting during readout in low gain mode if intstart is negative.)
  • Integrate

During the integration phase, the device's master clock is currently not operated. (This appears not to cause any adverse effects.)

The integration ends at the start of cycle n+1. The first cycle may also often request a "reset" rather than a readout for IR devices. For these reasons, nframes is often 1 or 2 greater than the actual number of desired video frames. (See the "keep" option for syntax to throw away data produced by these initial cycles.)

etime

The exposure time is specified in seconds. As of this writing, the controller pays attention to up to 3 decimal places (milliseconds) but a future upgrade will increase the precision to microseconds.

intstart

For devices with an electronic shutter, a time in seconds after the readout ends can be specified as the "start of integration." If it is desired to have integration overlap with readout, a negative number of seconds can be used (up to but not greater than the readout time, and also never greater than the etime.) No attempt is made to enforce the validity of integration during readout based on dreg. I.e., the user

width, height

These parameters default to the full (per output) detector size. If a window readout mode has been selected (see dreg) then width and height should be adjusted accordingly.

keep

When keep=false is used, no data will be transmitted. This is useful for a reset frame which should be discarded, or pipelined data which does not need to be kept. The default is keep=false. This option can be changed on a given frame (cycle) number using the extended syntax described under Frame-based options above.

mod

All of the frame based options can be set to repeat on a given number of cycles. For example, if "mod=5" is given, the 6th cycle will operate with the settings that were applied to the first cycle.

dreg

This option sets a device-specific data register. It can also be set to change on any cycle, but options which affect width and height (windowing) should remain the same.